Testing Performance 6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide 85
20 Repeat step 15 through step 19 to complete measuring gains for remaining
scale settings for channel 1 in Table 11 of the Performance Test Record.
21 For each of the remaining channels, move the splitter to the channel under test
and change the oscilloscope configuration as follows:
a Click Control > Factory Default.
b Ensure the channel under test is displayed and all other channels are turned
off.
c Set the vertical scale of the channel to 5 mV/div.
d Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz
waveform).
e Click Trigger> Setup Trigger... and change the source to the channel.
f Click Measure > Add Measurement.... Select the Vertical RMS voltage
measurement, the channel under test as the source, Entire Display as the
Measurement Area, and AC for the RMS Type.
g Repeat step 6d and step 7 to step 20 to complete measuring gains for the
channel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies Infiniium MXR404A and is the answer not in the manual?

Keysight Technologies Infiniium MXR404A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniium MXR404A
CategoryTest Equipment
LanguageEnglish

Related product manuals