Testing Performance 6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide 89
Performance Test Record
Model Number: Tested by:
Serial Number: Work Order No.:
Recommended Test Interval—1 Year/2000 hours Date:
Recommended next test date: Ambient temperature:
Table 5 Input Impedance Test, Channels 1 through 4
Input
Resistance
Vertical
Scale
Channel 1 Channel 2 Channel 3 Channel 4 Limits
50 5mV/div 48.25 to 51.75
50 10 mV/div 48.25 to 51.75
50 20 mV/div 48.25 to 51.75
50 50 mV/div 48.25 to 51.75
50 100 mV/div 48.25 to 51.75
50 200 mV/div 48.25 to 51.75
50 500 mV/div 48.25 to 51.75
50 1 V/div 48.25 to 51.75
1M 5mV/div 990k to 1.010 M
1M 10 mV/div 990 k to 1.010 M
1M 20 mV/div 990 k to 1.010 M
1M 50 mV/div 990 k to 1.010 M
1M 100 mV/div 990 k to 1.010 M
1M 200 mV/div 990 k to 1.010 M
1M 500 mV/div 990 k to 1.010 M
1M 1V/div 990k to 1.010 M
1M 2V/div 990k
to 1.010 M
1M 5V/div 990k to 1.010 M

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies Infiniium MXR404A and is the answer not in the manual?

Keysight Technologies Infiniium MXR404A Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniium MXR404A
CategoryTest Equipment
LanguageEnglish

Related product manuals