Testing Performance 6
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide 99
500 mV/div ±1.5 V ±2 %
1V/div ±3.0V ±2 %
Table 9 DC Gain Accuracy Test (50 ) (continued)
Vertical
Scale
Ext. Supply
Setting
V
DMM+
V
Scope+
V
DMM-
V
Scope-
Calculated
DC Gain
Error
DC Gain
Error Test
Limits
Table 10 DC Gain Accuracy Test (1 M)
Vertical
Scale
Ext. Supply
Setting
V
DMM+
V
Scope+
V
DMM-
V
Scope-
Calculated
DC Gain
Error
DC Gain
Error Test
Limits
Channel 1
5 mV/div ±15 mV ±2 %
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1V/div ±3.0V ±2 %
2V/div ±6.0V ±2 %
5 V/div ±15.0 V ±2 %
Channel 2
5 mV/div ±15 mV ±2 %
10 mV/div ±30 mV ±2 %
20 mV/div ±60 mV ±2 %
50 mV/div ±150 mV ±2 %
100 mV/div ±300 mV ±2 %
200 mV/div ±600 mV ±2 %
500 mV/div ±1.5 V ±2 %
1V/div ±3.0V ±2 %
2V/div ±6.0V ±2 %
5 V/div ±15.0 V ±2 %
Channel 3