Configure an External Current Probe
The IntegraVision PA2200 Series power analyzers support several types of current probes,
including external current shunts, current clamps, current transformers, Hall effect probes,
Rogowski coils, and the Keysight N2780B, N2781B, N2782B, and N2783B current probes.
These current probes typically produce a voltage output proportional to the measured current.
If your current probe produces a current output, you may route that output directly to an
internal current shunt on the rear panel, or you may route the current through an appropriate
resistor and use the voltage drop across the resistor as the input to the power analyzer's external
current input.
Wiring Diagram
A typical wiring diagram for an external current probe is shown below.
Keysight IntegraVision PA2200 Series Power Analyzers Operating and Service Guide41