Equations for three-phase wiring groups
In the following table, subscript numbers refer to the channel number in the wiring group. For
example, V
2
is the voltage on leg 2. The subscript "n" means "neutral." For example, V
1n
is the
voltage from leg 1 to neutral.
Equations in table are shown for AC+DC meas mode and calculate RMS using the formula:
∑
RMS x x=
N
n
N
n
1
=0
1
2
−
3P-3W-2M 3P-3W-3M 3P-4W-3M
Voltage as Wye
( )
V RMS v= −
n
v v
1 1
+
3
1 2
( )
V RMS v= −
n
v v
2 2
+
3
1 2
( )
V RMS v= −
n
v v
1 1
−
3
1 2
( )
V RMS v= − −
n
v v
3 2
−
3
1 2
( )
V V V V= + + / 3
avg n n n1 2 3
( )
V V V V= + + / 3
avg 1 2 3
Voltage as Delta
( )
V V V V= + + / 3
avg 12 23 31
Current
( )
I I I I= + + / 3
avg 1 2 3
I RMS i i i= ( + + )
n 1 2 3
Power
Keysight IntegraVision PA2200 Series Power Analyzers Operating and Service Guide119