129
Key Reference
Meas
Rel Stop
Couple Man
Enters a relative level limit at Stop Freq ranging from –150.00
to +50.00 dBc with 4 significant digits, and to toggle this
function between Couple and Man. If set to Couple, this is
coupled to Rel Start to make a flat limit line. If set to Man, Rel
Start and Rel Stop the user can enter different values to make
a sloped limit line.
Key access: Meas > SEM > Offset/Limits > More
Fail Mask Displays the menu to select one of the following logic keys
for fail conditions between the measurement results and the
test limits: Absolute, Relative, Abs AND Rel, and Abs OR Rel.
Key access: Meas > SEM > Offset/Limits > More > More
Absolute Sets the limit test to show Fail if one offset is
larger than the corresponding limit for Abs Start
and/or Abs Stop. This is the default selection for
each offset for BTS measurements.
Relative Sets the limit test to show Fail if one offsets
spectrum emission mask measurement results is
larger than the corresponding limit for Rel Start
and/or Rel Stop.
Abs AND Rel Set the limit test to show Fail if one offset’s
spectrum emission mask measurement results is
larger than both the corresponding limit for Abs
Start and Abs Stop and the limit for Rel Start and
Rel Stop.
Abs OR Rel Set the limit test to show Fail if one offset’s the
absolute spectrum emission mask measurement
results is larger than the corresponding limit for
Abs Start and Abs Stop or the limit for Rel Start and
Rel Stop.
Meas BW Specifies the bandwidth to use when measuring the offset.
When using the front panel, only the currently selected offset
is affected. All the offsets are affected when remote control.
This is the multiplier applied to the resolution bandwidth to
determine the bandwidth to integrate when calculating the
power. Meas BW is restricted to be 1 and not changeable.
Key access: Meas > SEM > Offset/Limits > More > More