119
Key Reference
Meas
Meas
In the Spectrum Analyzer mode (see MODE), Meas key displays
a menu for measurements such as adjacent channel power,
occupied bandwidth, channel power, spectrum emission mask
and TOI. Press Measure Off to turn the measurement off.
Channel Power
Channel Power measures the power and power spectral
density in the specified channel bandwidth. A white frame
on the display indicates the edge limits of the channel
bandwidth. The user must specify the center frequency,
reference level, and channel bandwidth to be used for the
measurement.
Center Freq Sets the center frequency of display.
Key access: Meas
> Channel Power
Integ BW Specifies the range of integration used in calculating the
power in a channel. For example, set the main (center)
channel bandwidth. Note that the white frame indicates the
integration bandwidth. Be sure the span is set between 1
and 10 times the integration bandwidth. Change integration
bandwidth by using the knob and the arrow keys.
Key access: Meas
> Channel Power
Span Sets the sweep span of display.
Key access: Meas
> Channel Power
Span Power Set the integrated bandwidth to the sweep span of display.
The channel power and power spectral density display on
the screen simultaneously.
Key access: Meas
> Channel Power