Key Reference
MODE
138
File Type
Toggles the saved file type between trace and CSV.
File Path
Allocate the path where the file is saved.
File Size
Set the number of frames for the spectrogram trace file. The
allowed input range is 200 to 1500 frames.
Reflection Measurement
The reflection measurement mode provides the reflection
measurement, one port cable loss and distance to fault
measurement function.
Auto Scale
Press this softkey to adjust the amplitude scale automatically
to include the peak and valley level in graticule.
Key access: Amplitude > Auto Scale
Calibrate
This softkey activates the calibration process for the
measurement.
Key access: Frequency > Calibrate
Cal Type
This softkey toggles the calibration type between full span
calibration and selected span calibration.
Full span calibration runs an open- short- load calibration
across the full frequency span.
A FullSpan calibration ignores the frequency changes during
measurements.
Selected span calibration performs an open- short- load
calibration across a user- defined frequency span. A
SelectedSpan Cal requires a re- calibration of the analyzer,