Key Reference
MODE
134
Meas Disp Accesses the Meas Disp submenu for defining the display
range and resolution while in Power Meter mode.
Key access: Mode > Power Meter
Zero Type Toggles the zeroing type of the Power Sensor
between internal and external. When this type is
set to EXT, pressing Zero will bring up a message
which indicates the external signal need be pulled
out. This setting is not affected by the preset in
Power Meter menu.
Key access: Mode > Power Meter > Zero/Cal
Calibration Toggles the zeroing type of the Power Sensor
between internal and external. When this type is
set to EXT, pressing Zero will bring up a message
which indicates the external signal need be pulled
out. This setting is not affected by the preset in
Power Meter menu.
Key access: Mode > Power Meter > Zero/Cal
Disp
Range
Accesses the associated Disp Range submenu.
Key access: Mode > Power Meter > Meas Disp
Auto
Range
Toggles the auto range function between
On and Off. When the Auto Range is set
to On, the Top and Bottom value of the
power meter’s display range will be set
automatically according to the
measurement data:
Top = Max. data + 0.5
Bottom = Min. data - 0.5
The default setting is Off.
Top Sets the top value of the power meter’
display range.
Range from (Bottom + 0.01) to 200 dBm.
The default value is 20 dBm.