EasyManua.ls Logo

Keysight N774 C Series

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N774xC Series Multiport Power Meter Programming Guide 101
Error Codes 5
-223 “Too much data”
Description:
A function returns more data or the user requests more data than the application is able to handle.
Example:
A tunable laser source produces more data when lambda values of a sweep are stored than the
816x instrument is able to handle. Use the new SENSE:FUNC:RES:BLOCK? command to split the
data aquisition into multiple parts.
-224 “Illegal parameter value”
[Used where exact value, from a list of possibles, was expected.]
-225 "Out of memory"
Description:
The request application or function cannot be executed because the instrument runs out of
memory.
-226 "Lists not same length"
-230 "Data corrupt or stale"
-231 "Data questionable (StatValNYetAcc)"
Description:
The data that is retured is not accurate or reliable. The user should repeat the operation. The
reason for this error is unspecific.
Example:
A powermeter configured a long average time has not completed its current measurement cycle
when the user queries the current power.
-231 "Data questionable (StatRangeTooLow)"
Description:
As -231 (StatValNYetAcc) but for a more specific reason: The powermeter readout data is not
reliable because the currently set (manual) range does not correspond with the input power.
-240 "Hardware error"
-241 "Hardware missing"
-250 "Mass storage error"
-251 "Missing mass storage"
-252 "Missing media"
-253 "Corrupt media"
-254 "Media full"
-255 "Directory full"
-256 "File name not found"
Error
Number String

Table of Contents

Other manuals for Keysight N774 C Series

Related product manuals