68 Keysight N774xC Series Multiport Power Meter Programming Guide
4 Measurement Operations & Settings
READ Subsystem Commands
This section provides the description of the following commands.
Command: :READ[n]:POWer:ALL?
Syntax: :READ[n]:POWer:ALL?
Description: Reads all available power channels. It provides its own software triggering and does not need a triggering command.
Parameters: None
Response: 4-byte Intel float values in a binary block in Intel byte order. The values are ordered by slot.
Data values are always in Watt.
Example: read:pow:all? -> interpreted as +1.33555600E-006|+1.34789100E-006|+1.37456900E-006
Affects: N7744C and N7745C
Command: :READ[n]:POWer:ALL:CSV?
Syntax: :READ[n]:POWer:ALL:CSV?
Description: Reads all available power channels. It provides its own software triggering and does not need a triggering command.
Parameters: None
Response: 4-byte Intel float values in a binary block in Intel byte order. The values are ordered by slot.
Data values are always in Watt.
Example: read:pow:all? -> interpreted as +1.33555600E-006|+1.34789100E-006|+1.37456900E-006
Affects: N7744C and N7745C