EasyManua.ls Logo

Keysight N774 C Series

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 Keysight N774xC Series Multiport Power Meter Programming Guide
4 Measurement Operations & Settings
CONFigure Subsystem Commands
This section provides the description of the following commands.
Command: :CONFigure:MEASurement:SETTing:ACTual?
Syntax: :CONFigure:MEASurement:SETTing:ACTual?
Description: Get the index of the setting currently being used.
Parameters: None
Response: A value >0 is returned if the setting has been stored in FLASH memory (using :CONFigure:MEASurement:SETTing:SAVE), or
has been recalled from FLASH memory (using :CONFigure:MEASurement:SETTing:RECall), and has not been changed since.
0 is returned if the setting has not yet been stored. 0 is returned if the FLASH setting has been deleted
(using:CONFigure:MEASurement:SETTing:ERASe) since the last recall or store. -1 is returned if the setting was changed but
has not been saved yet.
Example: :conf:meas:sett:act? +2
Affects: All instruments
Command: :CONFigure:MEASurement:SETTing:NUMBer?
Syntax: :CONFigure:MEASurement:SETTing:NUMBer?
Description: Get the number of settings. In addition to the settings spaces in FLASH memory, the working memory can hold a setting.
Parameters: None
Response: int
Example: :conf:meas:sett:numb? +1
Affects: All instruments
Command: :CONFigure:MEASurement:SETTing:PRESet
Syntax: :CONFigure:MEASurement:SETTing:PRESet
Description: Resets the setting values in the working memory. In contrast to the *RST and System:Preset commands, the previous stored
settings remain in nonvolatile RAM and can be recalled again.
Parameters: None

Table of Contents

Other manuals for Keysight N774 C Series

Related product manuals