Keysight N774xC Series Multiport Power Meter Programming Guide 39
Instrument Setup and Status 3
5 Not used Channel 5 Summary 32
4 Channel 4 Summary Channel 4 Summary 16
3 Channel 3 Summary Channel 3 Summary 8
2 Channel 2 Summary Channel 2 Summary 4
1 Channel 1 Summary Channel 1 Summary 2
0 Not used Not used 1
Example: :stat:ques:cond? -> +0
Command: :STATus:QUEStionable:ENABle
Syntax: :STATus:QUEStionable:ENABle<wsp><value>
Description: Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the contents of the QSESR to affect the
Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the QSESR to affect bit 3 of the Status Byte.
Parameters: The bit value for the questionable enable mask as a 16-bit unsigned integer value (0 .. +65535)
The default value is 65535.
Response: none
Example: :stat:ques:enab 128
Command: :STATus:QUEStionable:ENABle?
Syntax: :STATus:QUEStionable:ENABle?
Description: Returns the QSESM for the event register
Parameters: none
Response: The bit value for the QSEM as a 16-bit unsigned integer value (0 .. +65535)
Example: :stat:ques:enab?
-> +128