EasyManua.ls Logo

Keysight N774 C Series

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44 Keysight N774xC Series Multiport Power Meter Programming Guide
3 Instrument Setup and Status
Parameters: 24-hour time format: hours (0-23), minutes (0-59), seconds (0-59)
Response: none
Example: :syst:time 20,15,30
Command: :SYSTem:TIME?
Syntax: :SYSTem:TIME?
Description: Returns the instrument’s internal time.
Parameters: none
Response: The time in the format hour, minute, second. Hours are counted 0...23 (24 hour time format).
Example: :syst:time? -> +20,+15,+30
Command: :SYSTem:ERRor[:NEXT]?
Syntax: :SYSTem:ERRor[:NEXT]?
Description: Returns the next error from the error queue.
Parameters: none
Response: The number of the latest error, and its meaning.
Note: Every connection uses its own error queue
Example: :syst:err? -> -113,"Undefined header"
Command: :SYSTem:ERRor:COUNt?
Syntax: :SYSTem:ERRor:COUNt?
Description: Returns the total no. of errors.
Parameters: none
Response: The total count of errors.
Example: :syst:err:coun? -> 20