40 Keysight N774xC Series Multiport Power Meter Programming Guide
3 Instrument Setup and Status
Command: :STATus[n]:QUEStionable[:EVENt]?
Syntax: :STATus[n]:QUEStionable[:EVENt]?
Description: Returns the questionable status of slot n - the Questionable Slot Status Event Register (QSSER).
Parameters: none
Response: The results for the individual slot events (a 16-bit unsigned integer value, where 0 value 65535):
Bit
2-15
1
0
Mnemonic
Not used
Channel n: A Zeroing operation has failed
Not used
Decimal Value
2
Every nth bit is the summary of slot n.
Example: :stat0:oper? -> +0
Command: :STATus[n]:QUEStionable:CONDition?
Syntax: :STATus[n]:QUEStionable:CONDition?
Description: Returns the Questionable Slot Status Condition Register for slot n.
Parameters: none
Response: The results for the individual slot events (a 16-bit unsigned integer value, where 0 value 65535):
Bit
2-15
1
0
Mnemonic
Not used
Channel n: A Zeroing operation has failed
Not used
Decimal Value
2
Every nth bit is the summary of slot n.
Example: :stat0:ques:cond? -> +0