66 Keysight N774xC Series Multiport Power Meter Programming Guide
4 Measurement Operations & Settings
I
NITiate Subsystem Commands
This section provides the description of the following commands.
Command: :INITiate[n][:CHANnel[m]][:IMMediate]
Syntax: :INITiate[n][:CHANnel[m]][:IMMediate]
Description: Initiates the software trigger system and completes one full trigger cycle, that is, one measurement is made for selected [n].
In logging mode it triggers all channels independent from [n].
Parameters: None
Response: None
Example: init1:imm
init2:imm
Affects: N7744C and N7745C
Command: :INITiate[n][:CHANnel[m]]:CONTinuous
Syntax: :INITiate[n][:CHANnel[m]]:CONTinuous<wsp><boolean>
Description: Sets the software trigger system to continuous measurement mode.
Parameters: A boolean value:
0 or OFF: do not measure continuously
1 or ON: measure continuously
Response: None
Example: init2:cont 1
Affects: N7744C and N7745C