EasyManuals Logo
Home>Keysight>Measuring Instruments>N774 C Series

Keysight N774 C Series Programming Guide

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
Keysight N774xC Series Multiport Power Meter Programming Guide 41
Instrument Setup and Status 3
Command: :STATus[n]:QUEStionable:ENABle
Syntax: :STATus[n]:QUEStionable:ENABle<wsp><value>
Description: Sets the bits in the Questionable Slot Status Enable Mask (QSSEM) for slot n that enable the contents of the Questionable
Slot Status Register (QSSR) for slot n to affect the QSESR.
Setting a bit in this register to 1 enables the corresponding bit in the QSSER for slot n to affect bit n of the QSESR.
Parameters: The bit value for the QSSEM as a 16-bit unsigned integer value (0 .. +65535)
Response: none
Example: :stat0:ques:enab 128
Command: :STATus[n]:QUEStionable:ENABle?
Syntax: :STATus[n]:QUEStionable:ENABle?
Description: Returns the QSSEM for slot n
Parameters: none
Response: The bit value for the QSSEM as a 16-bit integer value (0 .. +65535)
Example: :stat0:ques:enab? -> +128