96 Keysight N774xC Series Multiport Power Meter Programming Guide
5 Error Codes
Error Strings
Error strings in the range -100 to -183 are defined by the SCPI standard,
downloadable from: http://www.ivifoundation.org/docs/scpi-99.pdf
Table 1 Overview for Supported Strings
Error
Number String
0 "No error"
-100 "Command Error"
[This is the generic syntax error used when a more specific error cannot be detected. This code
indicates only that a Command Error as defined in IEEE 488.2,11.5.1.1.4 has occurred.]
-101 "Invalid character"
[A syntactic element contains a character which is invalid for that type; for example, a header
containing an ampersand, SETUP. This error might be used in place of error -114 and perhaps
some others.]
-102 "Syntax error"
[An unrecognized command or data type was encountered; for example, a string was received
when the device does not accept strings.]
-103 "Invalid separator"
[The parser was expecting a separator and encountered an illegal character; for example, the
semicolon was omitted after a program message unit]
-104 "Data type error"
[The parser recognized a data element different than one allowed; for example,numeric or string
data was expected but block data was encountered.]
-105 "GET not allowed"
[A Group Execute Trigger was received within a program message (see IEEE488.2, 7.7).]
-108 "Parameter not allowed"
[More parameters were received than expected for the header]
-109 "Missing parameter"
[Fewer parameters were received than required for the header]
-110 "Command header error"
-111 "Header separator error"
-112 "Program mnemonic too long"
[The header contains more than twelve characters (see IEEE 488.2, 7.6.1.4.1).]