36 Keysight N774xC Series Multiport Power Meter Programming Guide
3 Instrument Setup and Status
Parameters: none
Response: The bit value for the operation enable mask as a 16-bit unsigned integer value (0 .. +65535)
Example: :stat:oper:enab? -> +128
Command: :STATus[n]:OPERation[:EVENt]?
Syntax: :STATus[n]:OPERation[:EVENt]?
Description: Returns the Operational Slot Status Event Register (OSSER) of slot n.
Parameters: none
Response: The results for the individual slot events (a 16-bit unsigned integer value, where 0
≤ value ≤ 65535):
Bit
4-15
3
1-2
Mnemonic
Not used
Channel n: Zeroing ongoing
Not used
Decimal Value
8
Example: :stat0:oper? -> +0
Command: :STATus[n]:OPERation:CONDition?
Syntax: :STATus[n]:OPERation:CONDition?
Description: Returns the Operational Slot Status Condition Register of slot n.
Parameters: none
Response: The results for the individual slot events (a 16-bit unsigned integer value, where 0
≤ value ≤ 65535):
Bit
4-15
3
1-2
Mnemonic
Not used
Channel n: Zeroing ongoing
Not used
Decimal Value
8
Example: :stat0:oper:cond? -> +0