EasyManua.ls Logo

Keysight N774 C Series

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N774xC Series Multiport Power Meter Programming Guide 47
Instrument Setup and Status 3
Command: :SYSTem:COMMunicate:ETHernet:AUTOip:ENABle
Syntax: :SYSTem:COMMunicate:ETHernet:AUTOip:ENABle
Description: Enable or disable whether IP addresses can be created automatically by the instrument. Automatic IP addressing is only
used if DHCP is enabled, but the instrument cannot find a DHCP server.
Parameters: Boolean (0 | 1 | off | on)
Response: None
Example: :SYST:COMM:ETH:AUTO:ENAB 1
Command: :SYSTem:COMMunicate:ETHernet:CANCel
Syntax: :SYSTem:COMMunicate:ETHernet:CANCel
Description: Undo all changes to the network parameters that have been made since the last save, reboot or ":syst:comm:eth:restart"
command.
Parameters: None
Response: None
Example: :SYST:COMM:ETH:CANC
Command: :SYSTem:COMMunicate:ETHernet:DGATeway
Syntax: :SYSTem:COMMunicate:ETHernet:DGATeway
Description: Set the default gateway.
Parameters: string (Up to four groups of up to 3 digits, groups separated by ".". Groups with leading zeros are interpreted as octal
numbers.)
Response: None
Example: :syst:comm:eth:dgat “192.168.101.11
Command: :SYSTem:COMMunicate:ETHernet:DGATeway?
Syntax: :SYSTem:COMMunicate:ETHernet:DGATeway?
Description: Get the default gateway.

Table of Contents