Keysight NFA Series Noise Figure Analyzers Service Guide     177
RF Section Troubleshooting (N8973B, 74B, 75B Analyzers)
Troubleshooting
Adjust the functioning spectrum analyzer to measure a signal at 4.8 GHz at 
+4.5 dBm ± 2 dB as shown in Figure 4-26.
Figure 4-26  4.8 GHz Signal
If this power level is incorrect, the most probable cause is the A16 Reference 
Assembly. If this level is correct and yet the signal previously measured at 
A14J740 is incorrect, the most probable cause is the A14 L.O. Synthesizer 
Assembly. Reconnect W4 cable to A14J740. Reconnect W6 cable to A14J200.