184   Keysight NFA Series Noise Figure Analyzers Service Guide
RF Section Troubleshooting (N8973B, 74B, 75B Analyzers)
Troubleshooting
6. Low Band Switch Control Logic Verification 
Press Mode Preset on the analyzer. Press FREQ, 50 MHz, Span, 2MHz on the 
analyzer. Connect the DVM positive lead to one of the In1A test points, and the 
negative lead to the bottom of R46 (blue resistor near the bottom) on the Front 
End Troubleshooting board. Verify the voltages in Table 4-13.
Press Freq 5 GHz on the analyzer. Verify the voltages in Table 4-14.
If the voltages are not correct, the most probable cause is the A15 Front End 
Control Assembly.
Table 4-13 
Test Point Voltage
In1A −9.85
In2A −9.85
In1B +9.75
In2B +9.75
Table 4-14 
Test Point Voltage
In1A +9.72
In2A −9.84
In1B −9.84
In2B +9.72