Keysight NFA Series Noise Figure Analyzers Service Guide 241
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Input Attenuator B Control Logic Verification (N8973B, 74B, 75B)
Press MODE/MEAS, Spectrum Analyzer, OK, AMPTD, Attenuation, Mech
Atten, 6 dB on the analyzer and verify the 6 dB Step LED DS17 is illuminated.
Change to the input attenuation settings found in Table 6-6 and verify the
proper LED's illuminate on the Front End Troubleshooting board according to
Table 6-6.
If the LED's illuminate correctly, the switch control logic for the input
attenuators from the A15 Front End Control Assembly is correct. If the LED's
are not illuminating as expected, the most probable cause is the A15 Front End
Control Assembly.
Set the input attenuator back to 10 dB by pressing AMPTD, Attenuation, Mech
Atten, 10 dB on the analyzer.
Table 6-6 Input Attenuator B LED’s
Attenuator
Setting (dB)
DS20
30 dB
Step LED
DS19
20 dB
Step LED
DS18
10 dB
Step LED
DS17
6 dB
Step LED
6 OFF OFF OFF ON
10 OFF OFF ON OFF
16 OFF OFF ON ON
20 OFF ON OFF OFF
30 ON OFF OFF OFF
40 ON OFF ON OFF
50 ON ON OFF OFF
60 ON ON ON OFF

Table of Contents

Other manuals for Keysight N8973B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N8973B and is the answer not in the manual?

Keysight N8973B Specifications

General IconGeneral
BrandKeysight
ModelN8973B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals