DOCUMENT 301900, REVISION D
EPISENSOR USER GUIDE 31
3. After checking that the O-ring is not pinched, screw the three large
seal screws through the cover and tighten to a torque of 18-20 inch
pounds (2.26-2.94 N.m).
4. Re-install the connector nut and tighten to a torque of 18-20 inch
pounds (2.26-2.94 N.m).
5. Re-install the two small seal screws in the X and Y zero-adjust
access holes on the side of the EpiSensor case.