- Start options with F2
In the parameter display, the text fields “Top temperature” and “Bottom temperature” are highlighted
in red.
F1: Main menu
You can access the screen for setting the threshold values on the supervisor level
The changes are effective immediately.
Page - 50 -
Here you can set a temperature window for each sealing jaw, which refers to the set target
temperature. If an actual temperature is outside the set tolerances, the device cannot be started.
To turn off the temperature alarm, set the limits to 0.0.
Seal seam strength tester Labormaster HCT 3000
Operating instructions and technical description
Overwrite the value with the number keys. Confirm your
choice with the ENTER key.
- System information with F6
5.3.20.2. Set temperature alarm
Use the TAB key to move the cursor (highlighted in blue) to the
limit value to be changed.
Picture:
Q2: The system information
F6: Startup options
Fax +49.7153.9510-100
Inh. Ludwig P. Goals
Ulmer Straße 51 – 53
E-Mail: sales@kopp-online.de
Figure 23 / Temperature alarm
D-73262 Reichenbach/Fils
Willi Kopp e. K
http://www.kopp-online.de
Phone +49.7153.9510-0
Packaging systems
Labormaster Version 2.6.3 SW 2.2 from 03/2016 Page 50 of 82, March 1, 2016
Subject to errors and technical changes!
Machine Translated by Google