The threshold value
is freely adjustable in the range of 1 ... 99%.
As a fixed threshold value in Newton: The
threshold value is
freely adjustable in the range of 2 ... 1700 N.
- Temperature alarm with F6
F6: Language
Use the TAB key to move the cursor (highlighted in blue) to the
threshold value to be changed.
When the sealing force reaches the threshold value, the sealing time expires.
Picture:
5.3.20.3. Startup options
5.3.20.3.1. Set sealing time threshold
F1: Main menu
F2: Temperaturalarm
You can enter the threshold value for the sealing time as an absolute or relative value.
- Language with F2
Seal seam strength tester Labormaster HCT 3000
Operating instructions and technical description
You can access the screen for setting the threshold values on the supervisor level
When the sealing force is reached, which corresponds to the percentage of the set sealing force, the sealing time
expires.
Page - 51 -
Factory setting: The starting
point of the sealing time is preset at the factory with a fixed threshold value of 60 N.
Move the cursor (highlighted in blue) to the selection menu using the
TAB key. Use the ÿ/ÿ arrow key to select the desired option
and confirm with the ENTER key
Note: If the threshold is higher than the sealing force achieved later during the test sealing, the sealing
time does not start. For safety reasons, the sealing rails open after a while and the error
message "Timeout threshold for sealing time not reached" appears in the lower edge of the display.
As a variable threshold in percent:
Overwrite the value with the number keys. Confirm your
choice with the ENTER key.
Switch between percentage and absolute information:
Packaging systems
D-73262 Reichenbach/Fils
Figure 24 / Start options
Fax +49.7153.9510-100
Phone +49.7153.9510-0
Inh. Ludwig P. Goals
E-Mail: sales@kopp-online.de
Willi Kopp e. K
http://www.kopp-online.de
Ulmer Straße 51 – 53
Labormaster Version 2.6.3 SW 2.2 from 03/2016 Page 51 of 82, March 1, 2016
Subject to errors and technical changes!
Machine Translated by Google