CardioCoach PRO User Manual | Page 74
5. Heart rate strap sensors will become worn out over time. A new heart
rate strap may be needed: https://korr.com/product/coded-heart-rate-
chest-strap/.
6. Bluetooth connectivity can be affected by a nearby Bluetooth devices.
Move away from other devices and/or shut off other devices to
reduce interference.
7. If Bluetooth signal is unreliable at your location, it may necessary for
you to use direct connection of heart rate pick up. KORR can provide
a Polar coded heart rate pick up from www.korr.com/coded-heart-
rate-pickup (KORR part number 9FG0190).
8. HR strap sensors will become worn out over time. A new HR strap
may be needed: https://korr.com/product/coded-heart-rate-chest-
strap/
9. Contact KORR Customer Support for further troubleshooting if
needed.
CardioCoach Indicates "Waiting" After Clicking "Start Test"
1. Click Abort.
2. Restart test.
3. If errors persist, contact KORR.
RMR Results that Seem Unusually High
Many obese people will test higher than you might expect. Studies
show that predictive formulas are particularly inaccurate for the
obese. When obese patients lift their heavy frame around, they
develop lean body mass simply to support their large bodies.
False high readings
o Clients breathing fast and shallow (hyperventilating)
o Clients moving around or fidgeting
o Clients who have eaten, consumed stimulants, or exercised
prior to testing
Research at the University of Utah indicates that the most helpful
environment for people to relax for an RMR test is to watch nature
videos with headphones on. The narration is distracting to the brain
while the subject matter fails to be stimulating.
RMR Results seem unusually low
Air leakage. Check for air leakage around the client’s mouth, at the
nose clip, or at the connecting point of the hose.
Re-test. A low metabolic rate may be a red flag of other medical
issues. If a second test is significantly low, refer to a physician for
follow up.