EasyManua.ls Logo

KORR CardioCoach PRO - VCO2 Specifications; Feco2 Specifications

KORR CardioCoach PRO
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CardioCoach PRO User Manual | Page 98
Minute Volume Specifications
Operating Characteristics
Accuracy
greater of ±5% or 0.3L
Resolution
0.1 L
Minimum Range
0 L
Maximum Range
100 L
Respiratory Rate Specifications
Operating Characteristics
Accuracy
2 breaths per minute
Resolution
1 breaths per minute
Minimum Range
4 breaths per minute
Maximum Range
30 breaths per minute
Resting Metabolic Rate (RMR) Specifications
Operating Characteristics
Accuracy
greater of ±10% of Reading or ±100 kcal/day
Resolution
7 kcal/day
Minimum Range
500 kcal/day
Maximum Range
5000 kcal/day
Measurement Time
10 min (typical)
VCO2 Specifications
Operating Characteristics
Accuracy
greater of ±10% Reading or ±10 ml/min CO
2
Resolution
1 ml/min CO
2
Minimum Range
70 ml/min CO
2
Maximum Range
4000 ml/min CO
2
FeCO2 Specifications
Operating Characteristics

Table of Contents