CardioCoach PRO User Manual | Page 95
Appendix
CardioCoach Specifications
The Absolute Maximum Ratings are those limits for the device that, if
exceeded, will likely damage the device. Exceeding the absolute maximum
ratings voids any warranty and/or guarantee. If the device is not used
properly, the protection offered by the equipment may be impaired.
The Operating Characteristics are those limits for the device that, if
exceeded, will cause the device to operate out of specification. Under these
conditions, values reported from the device will be unreliable. After the
“out-of-limit” condition is corrected, the device will return to normal
operation after either running an auto-calibration cycle or after the device
has been powered down and then powered back up.
Flow Port Specifications
Absolute Maximum Ratings
Maximum Flow Reading
300 LPM
Operating Characteristics
Type
Fixed Orifice
Accuracy
±2% of Reading or ±0.10 LPM
Extended Range Accuracy
±5% of Reading or ±0.50 LPM
Resolution
0.1 LPM
Minimum Range
-40 LPM (-667 ml/sec)
Maximum Range
300 LPM (5000 ml/sec)
Frequency Response
> 12 Hz
Sample Rate
60 Hz
Dynamic Resistance
< 2.0 cmH
2
O @ 60 LPM
Low Flow Drop Out
1.0 LPM
Type
Fixed Orifice