Installation and Settings
CardioCoach PRO User Manual | Page 92
9. Reconnect the power supply.
On the next powering on of CardioCoach the calibration data in the O
2
sensor will read into the system and be ready for use.
Cleaning and Storage
Recommendations for the Unit
• Protect the CardioCoach from extreme temperatures or moisture that
can damage internal components and sensors. Storage outside at 59 to
86°F (15 to 30
C) will cause damage and void warranty.
• Clean the exterior case with a lint-free damp cloth (70% Isopropyl
Alcohol). Do not use solvents.
Recommendations for Cleaning Accessories
The degree of risk of cross-infection of patients via cardio-pulmonary
testing equipment has yet to be quantified. Based on current evidence,
elaborate precautions are not justified, but attention to appropriate routine
cleaning and disinfection protocols is important. The following are
recommendations made based on the existing research as summarized in a
study published in Respiratory Medicine in 2003. (Infection control of lung
function equipment: a practical approach. A.H. Kendricka, D.P. Johnsb , J.P
Leeming. Respiratory Medicine (2003) 97, 1163-1179).
Masks and Y-valves
These valves, such as those from Hans Rudolph, are heat sensitive to
temperatures above 40 degrees C and are damaged by prolonged exposure
to hypocholorite solutions. Dismantling, washing in hand-hot soapy water,