MAL-OPS-1026 v3 70
ID_WRONG_HARDWARE
"Wrong
Hardware"
The device has had software
installed on it not intended for this
model. Device may no longer be
calibrated
ID_ILLEGAL_BUILD
"Unofficial
Firmware"
The device has had unauthorised
software installed on it. Device may
no longer be calibrated
ID_STORAGE_DRIVER
"Storage
Error"
An internal hardware error has
occurred, and device settings are
not accessible
ID_RPI_ERROR
"ID Module
Error"
An internal hardware error has
occurred, and the Nuclide ID
engine failed to start or has
stopped responding
ID_K40_CAL_ERROR
"Calibration
Error"
Dynamic K40 based calibration has
detected the device is out of
calibration
ID_K40_WARNING
"Calibration
Warning"
Dynamic K40 based calibration has
detected the device may need
recalibration
ID_NO_SDCARD "No SD Card"
The internal mass storage is not
installed or reporting as not
installed. N42 files are unavailable
ID_SDCARD_ERROR
"SD Card
Error"
The internal mass storage is
corrupt or not responding
ID_WIFI_INIT_ERROR
"Wi-Fi
Module
Error”
The internal Wi-Fi module is faulty
or not responding
ID_GNSS_I2C
"GPS Bus
Error"
The internal GNSS module is faulty
or not responding
ID_NO_GAMMA_COUNTS
“Gamma
Count
Timeout”
No counts were received from the
gamma detector. The detector may
be faulty
ID_NO_NEUTRON_COUNTS
"Neutron
Count
Timeout"
No counts were received from the
neutron detector. The detector
may be faulty
ID_NO_DOSE_COUNTS
"Dose Count
Timeout"
No counts were received from the
dose detector. The detector may
be faulty
ID_FRAM_CRC
“FRAM CRC
Failed”
Corruption has been detected on
the internal memory