Chapter - 5 Initial Startup / 5.2 Preparing the Instrument after Installation
23
Use the buttons and to select the desired
instrument settings (scale, polarimeter tube or
temperature).
Set the desired parameter with the button .
Exit the selected instrument setting with the buttons
and and select the next instrument settin
you want to change.
Exit the menu Instrument Settings
with
.
Determining the Blank Depending on whether the accuracy is checked using a
quartz standard or a standard solution, the blank
determination is performed differently as follows:
When performing the accuracy measurement with a
quartz standard, the blank measurement is carried
out with the sample chamber empty.
When performing the accuracy measurement with a
standard solution, the blank measurement is
carried out with an empty polarimeter tube (100
mm / 200 mm, selected in the instrument settings).
The determined blank is then subtracted from all
following measurement results.
Open the system settings with .
Select
using the buttons
and