Chapter - 5 Initial Startup / 5.2 Preparing the Instrument after Installation
25
Measuring the Standard
Insert the quartz standard into the sample
chamber.
or
Fill the standard into the polarimeter tube.
Close the sample chamber
Start the measurement with .
The measurement result is displayed.
If the result matches the value of the quartz standard or
the value of the standard solution, you may start with
routine operation.
If the results show larger deviations from the nominal
values, the measurement range should be re-
calibrated.
See 7.3 Calibrating the Measurement Range.