Lighthouse SOLAIR 3350, 5350 Gen E Operating Manual
7-6 248083387-1 Rev 10
The low word register for Sample Time is 40034.
The low word register for Hold Time is 40032.
The low word register for Initial Delay is 40030.
Table 7-4 Instrument Parameters
Register Data Type Description
40026 unsigned integer Location number; Specifies location of
Particle Counter.
40029 unsigned integer Initial Delay [high]. Works in conjunction
with 40030. Number of seconds to wait
before starting the first sample. Max value
is 86,399, which equals 23h 59m 59s.
40030 unsigned integer Initial Delay [low]
40031 unsigned integer Hold Time [high]. Works in conjunction
with 40032. Number of seconds to wait
between sample periods. Max value is
86,399, which equals 23h 59m 59s
40032 unsigned integer Hold Time [low]
40033 unsigned integer Sample Time [high]. Works in
conjunction with 40034. Number of
seconds to sample. Max value is 86,399,
which equals 23h 59m 59s.
40034 unsigned integer Sample Time [low]