LIGHTMED TruScan Pro Operator’s Manual Rev. No. 01 Page 100 of 109
No.
Warning Message
Description
Corrective Actions
1
“REMOTE DOOR
INTERLOCK ACTIVATED”
Treatment room door is
not closed
Remote interlock
connector is not
inserted
“Please close the
treatment room door”
“Insert the remote
interlock connector”
2
“FOOTSWITCH NOT
FOUND”
Footswitch is not
connected to laser console
“Please connect the
footswitch to the console
3
“FIBER PORT 1 NOT
FOUND
Slit lamp fiber is not
attached to Port 1
Endoprobe is not
attached to Port 1
“Please attach the slit lamp
fiber or endoprobe to port
1”
4
“FIBER PORT 2 NOT
FOUND”
LIO fiber is not attached to
Port 2
“Please attach the LIO fiber
to port 2”
5
“DELIVERY KEY NOT
FOUND”
Delivery key is not inserted
into the console
“Please insert the correct
delivery key into the
console”
6
“LASER DIODE 532 SHG
TEMPERATURE UNDER
REGULATION
Laser diode 532 second-
harmonic generation (SHG)
temperature is above or
below 3°C.
“Please wait while the
console adjusts itself”
7
“LASER DIODE 532 BRF
TEMPERATURE UNDER
REGULATION
Laser diode 532 BRF
temperature is above or
below 3°C.
8
“LASER DIODE 532 CASE
TEMPERATURE UNDER
REGULATION
Laser diode 532 case
temperature is above or
below 3°C.
9
“LASER DIODE 577 SHG
TEMPERATURE UNDER
REGULATION
Laser diode 577 SHG
temperature is above or
below 3°C.
10
“LASER DIODE 577 BRF
TEMPERATURE UNDER
REGULATION
Laser diode 577 BRF
temperature is above or
below 3°C.
11
“LASER DIODE 577 CASE
TEMPERATURE UNDER
REGULATION
Laser diode 577 case
temperature is above or
below 3°C.
12
“LASER DIODE 670 CASE
TEMPERATURE UNDER
REGULATION
Laser diode 670 case
temperature is above or
below 3°C.
13
“LASER DIODE 810 CASE
TEMPERATURE UNDER
REGULATION
Laser diode 810 case
temperature is above or
below 3°C.
14
“DO NOT PRESS THE
FOOTSWITCH”
User presses the footswitch
when the system is not in
“READY” mode
“Please release your foot
on the footswitch pedal or
check footswitch cable
connection”
15
“SHUTTER NOT READY”
Shutter not ready
“Please wait while the
console adjusts itself”
16
“COMMUNICATION
FAILURE”
Main program is wrong
“Please check the
console state”