LIGHTMED TruScan Pro Operator’s Manual Rev. No. 01 Page 102 of 109
above or below 10°C over
5 minutes
8
Err 5043
Laser diode 532 case
temperature regulation is
above or below 10°C over
5 minutes
9
Err 5044
Laser diode 577 SHG
temperature regulation is
above or below 10°C over
5 minutes
10
Err 5045
Laser diode 577 BRF
temperature regulation is
above or below 10°C over
5 minutes
11
Err 5046
Laser diode 577 case
temperature regulation is
above or below 10°C over
5 minutes
12
Err 5047
Laser diode 670 case
temperature regulation is
above or below 10°C over
5 minutes
13
Err 5048
Laser diode 810 case
temperature regulation is
above or below 10°C over
5 minutes
14
Err 5082
Flash memory error
7.4 3D Mouse Error Symptoms
If the 3D mouse is not able to carry out any actions during the normal operation of the laser
system, please refer to the list below:
No.
Error Symptoms
Description
Corrective Actions
1
Not working
3D mouse battery
is low, the LED
light is red
Charge the 3D mouse
2
No movements
on the LCD
screen
3D mouse is not
connected to the
system
Plug-in the 3D mouse and reboot the system
3
Shows circle
menu
3D mouse is not
configured
correctly
i. Click the left black button on the 3D
mouse, a menu will be prompted.