144/150 D9U001GKB-0101
Ingress protection
IPX4
Ingress protection of the foot controller
IPX6
Method of operation
Int.2/18 min
Ambient temperature
+5° to + 40°C
Loss of programmed functions - without power - after battery
discharge
after 10 minutes
Consumption in Standby mode (chair version with colposcope)
20VA (13W)
Consumption max. (chair version with sockets)
2000VA
Consumption max. (chair version without sockets)
290VA
Noise level
Less than 56 dB (A)
Frequency (100 V)
50 Hz
Frequency (120 V)
60 Hz
Frequency (230 V)
50/60 Hz
Battery operation:
If the battery capacity is below 50%, there is a beep after 2 seconds every time
the controls are used
If the system is activated and the plug is pulled, the system stops. In contrast, if
the system is running on batteries and the power cable is plugged in, the
system will continue operating.
The charging indicator light may blink when the system is operating with high
load causing a voltage drop and due to the fact that the battery begins to
charge.
CB20 with backup battery begins to charge only when connected to the mains.
Control box with battery should be recharged at least once every 6 months.
However, the maximum lifetime is reached if the battery is fully charged.
The battery should be charged at least 12 hours before use.