2.6 Setting-up the Chair
The chair Gracie shall be set up by the representatives of L I N E T spol. s r.o.
Any handling/maneuvers with the chair within the room is permitted according to
the instructions stated in the Section 4.
Before a patient gets on the chair Gracie, the chair must be set up
safely.
The chair is intended for patients with weight up to 180 kg, the total
load must not exceed 210 kg
The chair Gracie must not be used in areas intended for medical care that
can constitute a risk of fire or explosion due to air, oxygen or nitrous oxide in
combination with anesthetic or cleaning agents (such as operating theatre).
Distribution cable is included. It is forbidden to connect it through an
extension cable and it is forbidden to connect the product into common
multiple sockets.
No power cable of another device can lead over the product surfaces or over
a patient lying on it.
While handling the product, be careful not to pinch any other objects.
While handling and subsequently setting-up the product, consistently pay
attention to safe positioning and cabling outside the base. Careless setting-
up causing pinching of the electrical and control cables can result in
equipment damage or personal injury.
Prior to resetting the chair, check that no other persons are within reach of
moving parts.
Objects (furniture, stands, etc.), which could get into collision with the chair,
may not be placed in immediate proximity of the chair.
These operating instructions contain permitted accessories. Use of other
than original accessories approved in these operating instructions is not
permitted.
Prior to connecting the mains cable, check its condition; it must not have any
mechanical damage.
The power cable must not be hung up on the movable part of the chair. In
that case, any damage to the part of the chair or power cable can occur.
The chair includes firmware that can be updated only by an authorised
service technician.
This firmware is protected against unauthorised access by mechanical
housing (tool is needed to access), by seal (components with processor are
sealed), by exclusive compatibility with an authorised software tool and by
check of compatibility of the new firmware with the chair.
Objects (furniture, stands, etc.), which could get into collision with the chair,
may not be placed in immediate proximity of the chair. Front side of the chair
is set up on height-adjustable feet. In most cases, the chair can be fixed
without further manual adjusting. In case of uneven floor, when the chair
cannot be unintentionally set up (it swings), the chair needs to be inclined to
the side and, using tools, adjusted to required height of the front foot for
secure set up. Then the chair needs to be inclined back to the base. The
chair may not swing!
A battery is a component of every gynaecologic chair, it is fully charged and
flawless and allows for the operation of the chair for the time of approx. 3
minutes. The battery is recharged automatically from the control unit and
does not require any special maintenance. If the chair is not connected to