• Portable and mobile HF communication devices should not be used
closer than the recommended distance of d = 15 cm from any part of
the chair, including the cables.
• The gynaecological chair, colposcope and monitor must be installed
and operated in accordance with the EMC information by
professionally trained service personnel.
Portable and mobile RF communications equipment can affect medical
electrical equipment.
Manual and declaration of the manufacturer - electromagnetic radiation
The chair is intended for use in electromagnetic environment specified below.
The customer or user of the chair must ensure that the chair is used in such
environment.
High frequency
radiation
CISPR 11
The chair uses high frequency energy
only for its internal functioning.
Therefore, its high frequency
radiation is very low and is not likely
to cause any interference with nearby
electronic equipment.
High frequency
radiation
CISPR 11
The chair is intended for use in all
facilities other than households and
buildings which are directly
connected to a public low-voltage
power supply network that supplies
buildings used for residential
purposes.
Harmonic
emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3