3) Displaying
Verification Scale Interval is 0.5 kg.
Press button 2 to display value with actual scale interval 0,1 kg for 5 s.
Display shows normally actual weight if other functions are not activated.
4) Hold Mode
Hold Mode must be used only when the scales are stabilized.
It allows adding or removing bed accessories and other items without changing the weight value.
To activate Hold Mode:
Wait 5 s until the scales are stabilized. The LED 10 will be illuminated when the scales are stabilized.
Press button 5 for 2 s.
The display shows „HOLD“, indicator of activated Hold Mode is illuminated.
Add or remove required accessories.
To deactivate Hold Mode:
After adding or removing accessories wait 5 s, until the scales are stabilized on the display.
The LED 10 will be illuminated when the scales are stabilized.
Press button 5 for 2 s.
The display shows the original weight value and the indicator of Hold Mode is not illuminated.
To deactivate Hold Mode without fixing the weight value:
Press button 4.
5) Setting Mode
To set date, date format, time and unit of weight:
Press button 4 and button 5 simultaneously for 3 s.
The value to be changed flashes on the display.
To navigate in list:
Press button 4 or button 5 to go up or down in the following list:
1.
minutes
2.
hours
3.
date format (month-day/day-month)
4.
year
5.
month
6.
day
7.
unit of weight
To leave setting mode:
Press button 4.
Setting Mode is left without saving the last setting.