! Important ! Adjusting light intensity levels in the COUNT screen will only be applied
to the live view mode. Exposure levels for brightfield cell counting are automatically
adjusted. Exposure levels for fluorescence cell counting may only be adjusted within
a protocol (Section 5).
Focusing
To bring cells into focus, press Autofocus. Focus may also be adjusted manually using the
up & down arrows in the focus control bar.
Zoom
Zoom in or out by spreading or pinching two fingers. The outer box in the upper, left hand
corner of the viewing window represents the current field of view. The inner box represents
the view on the screen. Zooming in or out will cause the inner box size to decrease or increase.
Navigation
To view different slide chambers within a slide, select the chamber to be viewed by pressing
a chamber on the slide image just under the COUNT button. To see different fields of view
within a chamber, use the arrows located above the manual focus adjustment.
field is in focus, press the COUNT button.
The LUNA-FX7
™
will count all slide chambers as designated in SETTINGS.
Counting time will vary depending on slide type, counting mode and the protocol used.