27
Protocol Parameters
Brightfield Cell
Counting
Parameters
Parameters for [Brightfield cell counting-Total cell counting]
Parameter
Range
DEFAULT*
Min. search size (μm)
1-89
7
Max. search size (μm)
2- 90
24
Cell detection sensitivity
1-10
5
Noise reduction
0-9
5
Dilution factor
1-100
1
Parameters for [Brightfield cell counting-Cell counting & Viability]
Parameter
Range
DEFAULT*
Min. search size (μm)
1-89
7
Max. search size (μm)
2- 90
24
Cell detection sensitivity
1-10
5
Live cell sensitivity
1-10
5
Noise reduction
0-9
5
Dilution factor
1-100
2
The IQOQ-BF on the Protocol List is the protocol that is used for IQOQ with the brightfiled
validation slide. Protocol YEAST is an optimized protocol that is used for yeast cell counting.