38
Press GRAPH to view a graphical representation. Press the Y-axis title box to s
witch between
‘Concentration’ and ‘Viability. Press ‘Day’, ‘Month’ or ‘Year’ to alter X-axis scale.
Press EXPORT to export a CSV file with count data and graph images to a USB drive.
Press RESULTS to view the most recent QC count.
! Important ! The value of total concentration printed on the slide label may differ from
the counting result of using the default protocol because the label value is determined
with a different protocol, which is for the purpose of IQOQ and Quality Control mode.
! Important ! After the Quality Control counting is done, remove the validation slide
from the device right away and store it in the case provided. The fluorescence pre-
fixed beads may be photobleaching if they are exposed to light.