EasyManua.ls Logo

logos biosystems LUNA FX7 - Page 48

logos biosystems LUNA FX7
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
Calibration
The LUNA-FX7 is calibrated prior to shipping. Calibration only needs to be performed 1)
after any software/firmware updates, and 2) after switching Trypan blue or Erythrosin B or
Methylene blue brands or formulations.
To perform cal ibration:
1. Press S/W & CALIBRATION.
2. Press Start calibration. The Calibration Step 1 window will appear.
3. Remove any counting slide from slide port.
4. Press START.
5. Press START or press SKIP if only fluorescence calibration is needed.
6. The Calibration Step 2 window will appear upon completion.
7. Load 10 μL of the LUNA-FXGF Calibration Beads into chamber A and 10 μL of the LUNA-
FXRF Calibration Beads into chamber B of a new PhotonSlide™. Vortex the calibration
beads solution thoroughly before loading into the PhotonSlide.
8. Wait for 30 seconds for the beads to settle down.
9. Insert the slide face up and chamber A first into the slide port.
10. Press START; or, press SKIP, if fluorescence calibration is not needed.
11. Upon completion, the calibration value(s) and calibration date will be updated.

Table of Contents

Related product manuals