PyroTec™ PRO Hardw are User’s Manual, V1.2 26
RT-MN024-1.2 02/21
Label
The identification data is also printed on the serial number label (B). This label can be
viewed from the front side of the instrument and is attached below the diluters.
Heights
Instrument equipped with:
Air displacement pipetting arm, Air LiHa
Robotic manipulator arm, RoMa
Safety Panel
Opening
Dimensions
Dimensions of the opening in the front safety panel:
PyroTec™ PRO: 1130 x 170 mm (44.5 x 6.7 in.)
Line voltage (single phase)
100 - 120, 220 - 240 V AC
(-15% / +10%)
2 x T10A (instrument power)
• x T2A (main powered op-
Electrical
Safety
Note: Considering possible under voltage on the mains supply, the combined power con-
sumption in the lower input voltage band (100 - 120 VAC) must not exceed 1000 VA in or-
der to keep the input current below the fused values.
Classification with regard to electrical safety according to EN/IEC standards:
The power switch is placed at the level of the front access panel. The power switch does
not switch the mains voltage directly, but gives a control signal to the power supply.
By unplugging the instrument.