PyroTec™ PRO Hardw are User’s Manual, V1.2 27
RT-MN024-1.2 02/21
At installation or later movement of the instrument, ensure that it is always possible to un-
plug the mains cable at the instrument.
ble Power
Supply (UPS)
For an optimal operation of instruments and smooth running of the relevant application,
Lonza recommends connecting an online UPS, so that the power supply runs via the UPS
with a filter effect.
UPS recommendations:
1.5 times the average consumption of the equipment.
Recommended minimum UPS
size
Note: A switched UPS type that switches over to batteries only after network breakdown is
not recommended.
For further assistance, contact your site manager or your nearest Lonza representative.
A switched UPS type that switches over to batteries only after network breakdown is not
recommended.
The status lamp is located above the diluters in the center of the instrument front surface.
It displays the instrument operational states with red and green light that is either contin-
uous or flashing. When the status lamp light is red, an acoustic alarm sounds.
The illuminated area is 540 x 18 mm (21.26 x 0.7 in.).
Conditions
All instruments are intended for indoor operation and storage only. The tables below give
an overview.
Barcodes cannot be read due to the influence of sunlight or other light sources on the bar-
code scanner.
• Do not expose the instrument to direct sunlight.
• Do not install strong light sources that may impair the function of the barcode scanner
near the instrument.
Conditions
15°C to 32°C (59°F to 90°F)
30% to 80% relative (non-condensing) at 30°C (86°F) or be-
low
Max. 2000 m above sea level