Page 112 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
9.7. PRINT FLOW ADJUSTMENT DATA
Note
Please consult the MAS-100 Iso NT Installation Manual for detailed information
about the printer installation.
PROCEDURE
1) Connect a printer to the air sampler.
2) Select in the pull-down menu the adjustment to
print:
Current adjustment
Last adjustment
3) Press Print adjustment data.
The data can be exported to a PDF file, if a virtual
PDF printer is available in the Windows® print
menu.
EXAMPLE - ADJUSTMENT DATA