10.7. PRINT FLOW ADJUSTMENT DATA
Note
Please consult the MAS-100 Iso NT Installation Manual for detailed information
about the printer installation.
PROCEDURE
1) Connect a printer to the air sampler.
2) Start up the air sampler by connecting it to the power.
While starting up, keep the buttons ❶ and ❹ pressed simultaneously for approxi-
mately 3 seconds.
3) The air sampler is now in the Fabtest menu.
Choose Adj./Cal. Mode,
then press Select.
4) The air sampler is now in the Adj./Cal. Mode menu.
Choose Print Cal. Data,
then press Select
Pressing Exit reverts to the Main menu.
5) Press OK.
Pressing Back interrupts the process and reverts to
the Adj./Cal. Mode menu.
6) The data is printed.
7) The air sampler reverts to the Adj./Cal. Mode menu.
8) End of process.