User Manual MAS-100 Regulus en.docx, Version: 4.0
5.5. AUTOMATIC ADJUSTMENT + CALIBRATION WITH CERTIFICATE
PROCEDURE
1) Perform the following steps in a clean and draft-free place, with all instruments placed
on a stable base.
2) Remove the perforated lid.
3) Check the lower edge of the perforated lid visually
to ensure that there are no notches or unevenness.
This could lead to a leak between the perforated lid
and the base.
Damaged perforated lids must be replaced.
4) Remount the perforated lid.
5) Place the anemometer on the perforated lid of the air sampler.
6) Connect the air sampler to the anemometer with the suitable communication cable.
7) Connect the air sampler to the PC with the suitable communication cable
8) Start the anemometer by pressing the ON/OFF
button.
The anemometer is in mass flow mode, showing
SLPM on the display.
9) Start the air sampler.
10) Start the PC.
11) Start the PC Software and login:
Access level: Calibrator
Password: mascal
12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
Base
Lid