User Manual MAS-100 Regulus en.docx, Version: 4.0
13) Select the perforated lid of the sampling head:
300×0.6 or 400×0.7
14) Select Automatic adjustment.
15) Set the adjustment parameters:
Adjustment Flow
Adjustment Span
Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the adjustment Flow is 100SLPM.
Note
The standard factory setting for the adjustment Span is 10SLPM.
16) Press Yes to proceed.
Do not add or remove instruments.
Pressing No aborts the process.
17) The air sampler will now adjust through all three adjustment points. No action from
the user is required, the process will continue automatically.
18) Press the Yes button to accept the adjustment and
store the values in the air sampler.
Pressing No aborts the process.
19) If this process must be repeated for different perforated lids, start over on step 12 and
use the other perforated lid.
20) End of process.